The function 1 chromatogram is not visible in MassLynx chromatogram view - WKB44898

Article number: 44898

SYMPTOMS

  • Both acquisition function panes are visible in MassLynx chromatogram view, but only the function 2 chromatogram is seen
  • The Show All Functions option is selected at Tools > Options > Chromatograms to display
  • The Add/Replace Chromatogram toggle button is unselected with Chromatogram view
  • The Reset display range toggle button is selected

ENVIRONMENT

  • MassLynx

CAUSE

The default graph view in MassLynx is set to "current".

FIX or WORKAROUND

In MassLynx chromatogram view, go to Display > Range > Default and set the default graph from "current" to "All".

ADDITIONAL INFORMATION