Cannot use Direct Injection in MassLynx for 3767 Sample Manager - WKB343654

Article number: 343654

SYMPTOMS

  • There is no option to enable Direct Injection in MassLynx for the 3767 Sample Manager
  • Not seen from SCN 1014 up to and including SCN 1051
  • Illustration below shows where the option used to be in the 3767 injector editor 
    Direct Injection 3767.png

ENVIRONMENT

  • 3767 Sample Manager
  • Autopurification
  • FractionLynx
  • MassLynx

CAUSE

Functionality was removed in SCN 1014 and in subsequent MassLynx builds up to and including SCN 1051.
Recorded in CRI-9928 Enable Direct Injection Functionality on 3767 in MassLynx.

FIX or WORKAROUND

Direct Inject functionality will be enabled in SCN 1054, which is due for release end of Q1 2027.