Microsoft Windows Defender keeps switching on - WKB329302

Article number: 329302

SYMPTOMS

  • Windows Defender cannot be disabled; on a newly imaged Waters PC, Windows Defender software is enabled
  • Attempting to disable the software by registry keys or other hacks does not work; after roughly 10 minutes, the registry keys disappear and Defender is active again
  • This can cause issues like the computer running very slowly when connected to the company network.
  • MassLynx and processes like Inlet Editor and Console may be very slow to respond. 

ENVIRONMENT

  • Windows 11
  • Windows 10
  • Empower
  • MassLynx
  • waters_connect

CAUSE

Microsoft has made a change to Windows so that Defender antivirus software is on by default. It cannot be disabled. See this Microsoft article: 

Microsoft Defender Antivirus compatibility with other security products

Defender will be deactivated by Microsoft ONLY when a non-Microsoft antivirus product is installed.

FIX or WORKAROUND

  1. Install a non-Microsoft antivirus product.
  2. Ensure that the correct exclusions are configured. See Recommended antivirus exclusions, firewall setup and UAC settings for MassLynx - WKB21150

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.