PDA detector does not start acquiring upon sample injection - WKB30552
Article number: 30552
SYMPTOMS
- PDA detector does not start acquiring data upon injection.
- MassLynx reloaded on new PC.
- ZQ2000 starts data acquisition as expected.
- Found: Under Inlet Method > Tools > Instrument Configuration > Events and Trigger, the page to set Trigger by Contact Closure or Software is missing.
- Found: Inlet configured with ACQUITY 2996 detector.
ENVIRONMENT
- MassLynx 4.1
- 2996 PDA Detector
CAUSE
Incorrect detector selected in the Inlet Configuration.
FIX or WORKAROUND
- Open Inlet Method.
- Go to Tools > Instrument Configuration.
- Click Configure.
- Step through the wizard to Configure Detectors page.
- Remove ACQUITY 2996.
- Add Waters 996 PDA.
- Complete the Wizard.
- Click Finish to close the Instrument Configuration dialog.
ADDITIONAL INFORMATION
id30552, 2996, MLYNX, MLYNXV41, SUPMM