No PDA trace in MassLynx when running samples - WKB291829
Article number: 291829
SYMPTOMS
- No PDA trace in MassLynx when running samples
- Inlet Editor Status OK indicator is red with text "No information currently available"
ENVIRONMENT
- ACQUITY PDA Detector
- MassLynx
CAUSE
Loss of communications with PDA and software
FIX or WORKAROUND
- Close MassLynx and associated windows.
- Run StopMassLynxProcesses.
- Open Waters DHCP Server Configuration.exe from C:/Program Files (x86)/Waters Instruments folder.
- Remove the PDA from the DHCP server configuration.
- Power-cycle the PDA and ensure that the PDA repopulates in the DHCP server configuration.
- Start MassLynx and confirm that communications are restored.