SYMPTOMS

  • After upgrading to SCN 1046, display issue seen:
  • When fractions are observed in the Chromatogram window at original size, the fraction IDs appear correct but are challenging to decipherFractionLynx_Chromatogram_Normal.png
  • When expanding the Chromatogram plot to examine fractions, they do not display correctly
  • Zooming in, the assigned fraction numbers shift arbitrarily and are no longer aligned with their fraction
  • Fraction colors are ignored, and the entire plot might be one color instead of multiple colours segregated by fractionFractionLynx_Chromatogram_Zoom.png

ENVIRONMENT

  • FractionLynx
  • Autopurification
  • MassLynx

CAUSE

Software bug - CRI-7273

FIX or WORKAROUND

Fixed in SCN 1049 

Workaround - view the data on a standalone PC with an earlier version of MassLynx. 