Injections do not run in MassLynx; sample list starts but the system never advances to the next injection - WKB281015

Article number: 281015

SYMPTOMS

  • Injections do not run in MassLynx
  • Sample list starts but the system never advances to the next injection or shows any progress
  • Deleting the queue and restarting has no effect
  • MassLynx shows the instrument is present
  • Resetting the LC or rebooting the PC has no effect

ENVIRONMENT

  • MassLynx

CAUSE

Event 1 is not set in Events & Triggering.

FIX or WORKAROUND

  1. Open the Inlet Method Editor.
  2. Open the Inlet Configuration menu:001 MassLynx Inlet Configuration.jpeg
  3. Open Events & Triggering as shown above.
  4. When you are in the Events & Triggering menu, click NEXT:clipboard_ee1a04c4efab79cb49c16f02ff841ba8c.png
  5. Clicking Next brings you to this screen:clipboard_e4de7b6f2a11e1fdf7e80e0a0c7a708f6.png
  6. Ensure that Event In 1 is selected.
  7. Advance through the rest of the screens, pressing NEXT until you are able to Finish and exit the Events & Triggering menu.