Skip to main content
Waters

There are no values in the Standard Conc column of the summary bar in the TargetLynx browser - WKB27089

Article number: 27089

SYMPTOMS

  • No values in the Std Conc column in the TargetLynx summary bar
  • No values for the y axis of the calibration curve
  • Standard concentrations are entered in the Conc A column in the sample list
  • TargetLynx processing method has Conc A in the Concentration of standard:level field
  • Select Calibrate standards in the Create TargetLynx XS Dataset window
  • Peaks integrated properly

ENVIRONMENT

  • MassLynx
  • TargetLynx

CAUSE

The Use Quan Ion in Response Calculation field is set to No. Change the field to Yes

FIX or WORKAROUND

In the TargetLynx processing method:

  1. Click the Target Ion Properties icon.
  2. Set Use Quan Ion in Response Calculation to Yes.
  3. Save the method.
  4. Reprocess the data.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.