There are no values in the Standard Conc column of the summary bar in the TargetLynx browser - WKB27089

Article number: 27089

SYMPTOMS

  • No values in the Std Conc column in the TargetLynx summary bar
  • No values for the y axis of the calibration curve
  • Standard concentrations are entered in the Conc A column in the sample list
  • TargetLynx processing method has Conc A in the Concentration of standard:level field
  • Select Calibrate standards in the Create TargetLynx XS Dataset window
  • Peaks integrated properly

ENVIRONMENT

  • MassLynx
  • TargetLynx

CAUSE

The Use Quan Ion in Response Calculation field is set to No. Change the field to Yes

FIX or WORKAROUND

In the TargetLynx processing method:

  1. Click the Target Ion Properties icon.
  2. Set Use Quan Ion in Response Calculation to Yes.
  3. Save the method.
  4. Reprocess the data.

ADDITIONAL INFORMATION

 

id27089, MLYNX, MLYNXV41, SUPMM, TARLYNX41