What are the required rights in MassLynx Security to allow for the import of projects to create checksum? - WKB27085
Article number: 27085
ENVIRONMENT
- MassLynx 4.1
- MassLynx 4.2
- MassLynx Security
ANSWER
In the Manage Groups > Assign rights, the user needs to have General > Import Legacy Project.
In the Directory Policies, assigned directory paths must have Import enabled.
ADDITIONAL INFORMATION
Note: The MassLynx user specifying their signature credentials for the Import of a project or file must be the user who logged into MassLynx. If Import: Legacy Project is enabled under the Dual Authorization Policy, then the Import action can only be performed by a User. It cannot be performed by an Administrator.
id27085, MLYNX, MLYNXV41, SUPMM