How to acquire only PDA data on an LC/MS system controlled by MassLynx - WKB2415
Article number: 2415
OBJECTIVE or GOAL
Acquire only the PDA data on an LC/MS system controlled by MassLynx.
ENVIRONMENT
- MassLynx 4.1
- Quattro Series
- ACQUITY SQ/TQ/3100
- Xevo Series
- SYNAPT Series
- ZQ/EMD Series
PROCEDURE
- Create a "dummy" MS channel as follows:
- Create an inlet method to acquire the appropriate PDA.
- Create a "dummy" MS Tune file with minimal allowed source temperature, desolvation temperature and desolvation gas flow rate
- Create a "dummy" MS method with a run time equivalent to that of the inlet method.
- In the "dummy" MS method, add a simple function to collect minimal data (example: single SIR or MRM transition with 0.1 s dwell time)
- In the MS method, go to Options > Method Events, under Initial Settings, set the Flow State to Waste to divert the flow away from the probe.
- Enter the inlet method, "dummy" MS method and "dummy" MS Tune file into their appropriate columns on the MassLynx Sample list.
Technically, this setup will still acquire MS data. In a regulatory environment, you may need to have no MS data channel at all. In this case, do the following:
Set the "dummy" MS method to acquire a minimal amount of data. Example: Single SIR with dwell time of 1 second, or very high data threshold or mass range that is out of the sample scope.
ADDITIONAL INFORMATION
NOTE: This workflow applies to other detectors such as the TUV, Fluorescence, ELSD, etc.
There are two additional alternatives, but are more "invasive" and more permanent solutions.
- Reinstall MassLynx with the option "No MS instrument" selected.
- Or, use two MassLynx PCs—one with an MS instrument selected and one with no MS instrument—and switch between them depending on the analysis.
