ACQUITY system under-pressure and over-pressure errors - WKB241380

Article number: 241380

SYMPTOMS

  • When trying to load method, system shows over-pressure errors at 7500 psi
  • When attempting to troubleshoot the error, system shows under-pressure error at 1000 psi
  • Both errors appear in the span of a few minutes

ENVIRONMENT

  • ACQUITY UPLC
  • ACQUITY I-Class

CAUSE

The inlet method high and low pressure limits were altered.

FIX or WORKAROUND

Reset the high and low pressure limits back to their default values: Low pressure limit of zero; high pressure limit of 15000 psi on ACQUITY UPLC, 18000 psi on ACQUITY I-Class.

ADDITIONAL INFORMATION