Error in MassLynx when submitting samples to run: Scan speed in function 1 is outside the calibrated range of 100 to 2000 amu per second. Do you want to apply the calibration? - WKB235805

Article number: 235805

SYMPTOMS

  • "Scan speed in function 1 is outside the calibrated range of 100 to 2000 amu per second do you want to apply the calibration"
  • Customer scan range is 60 to 2034, scan time is 0.3 sec, scan speed is 3253 amu/sec
  • Customer is running an MS Scan

ENVIRONMENT

  • ACQUITY TQ Detector
  • Xevo TQ-GC
  • MassLynx

CAUSE

The scan speed, based on setting in the MS method, is too high.

FIX or WORKAROUND

There are two fixes:

  1. In the MS method, set the Scan Time to 1 second. This will bring the scan speed to within the calibrated scan speed range.
  2. In the MS method, change the function to MS2 scan (Function > MS2 Scan), leaving the scan range and scan time the same.

ADDITIONAL INFORMATION

On the ACQUITY TQD, MS1 is calibrated from 100 to 2000 amu/sec.  MS2 is calibrated from 100 to 10000 amu/sec.