Using no calibration curve, how do you calculate the concentration in TargetLynx from the concentration of the internal standards? - WKB234165
Article number: 234165
OBJECTIVE or GOAL
Calculate the concentration of the analyte from its internal standard using no calibration curve.
ENVIRONMENT
- MassLynx
- QuanLynx
- TargetLynx
PROCEDURE
- Create the component for the Internal Standard in the TargetLynx method.
- Specify the concentration of the Internal Standard on the Calibration Properties page in the TargetLynx Method, in either of two ways:
- Set a Concentration of Standard: Level as Fixed and specify Concentration of Standard to that of the Internal Standard.
- Set a Concentration of Standard: Level to the Conc X column in the sample list. Specify the Concentration of the Internal Standards in this column.
- Create the component for the Analyte in the TargetLynx method.
- On the Compound Properties page, set the Response Type to Internal (relative - use internal standard).
- Set Internal Standard 1 to the appropriate Internal Standard from the Component List.
- On the Calibration Properties page, set the Polynomial Type to Manual RF.
- On the Calibration Properties page, set the User RF Value to 1.0.
- Save the method.
ADDITIONAL INFORMATION
The calculation for Response when using Response Type = Internal (relative - use internal standard) is:
Response = (Area Analyte / Area IS) x Concentration IS
Using a Response Factor (RF) of 1.0 converts this Response value directly to a Concentration value.
id234165, MLYNX, MLYNXV41, SUPMM, TARLYNX41