How to determine whether the APCI was run in Current mode or Voltage mode - WKB231877

Article number: 231877

OBJECTIVE or GOAL

  • Determine whether the APCI was run in Current mode or Voltage mode.

ENVIRONMENT

  • MassLynx
  • TargetLynx

PROCEDURE

  1. Open the chromatogram or TargetLynx Experiment for the data of interest.
  2. Instrument Parameters - Function X:

Parameter File - C:\MassLynx\Waters_2021_Datousei.PRO\ACQUDB\Phenol2.ipr
Polarity    API-
Calibration    Static 2
Corona (µA)    0.30
Corona (kV)    2.5

The set values for both modes are displayed here.

3. Check the instrument readbacks:


Corona (µA) 23.09
Corona (kV) 2.5

In this case, it is Corona (kV) 2.5, so it was run in Voltage mode.