Error: "There is no email program associated to perform the requested action" when opening OAManager with Thunderbird configured - WKB225470

Article number: 225470

SYMPTOMS

  • A  Mozilla Thunderbird email account is configured on the PC
  • Opening OAManager gives the following message:
    "There is no email program associated to perform the requested action. 
     Please install an email program."
  • The email account is correctly configured and set as the default email program
  • Other systems with the same email program do not show the error
  • Emails still send correctly

ENVIRONMENT

  • MassLynx
  • OALogin
  • OAManager
  • Mozilla Thunderbird email

CAUSE

A recent update to Thunderbird is causing issues with the email being recognised

FIX or WORKAROUND

  1. Downgrade to an earlier version of Thunderbird or use a different email system
  2. Or click away the error message and continue. Emails still sent.

ADDITIONAL INFORMATION

 