No LockSpray signal in the lockmass function of an LC-MS analysis - WKB219495
Article number: 219495
SYMPTOMS
- No spray signal in the lock mass function of an LC-MS analysis
- LockSpray signal good in the tune page
ENVIRONMENT
- MassLynx
- SYNAPT G2-Si
CAUSE
Method events disabled or wrong or LockSpray properties wrong
FIX or WORKAROUND
- Open the MS method
- Enable method events and select the correct events or select the correct LockSpray properties
ADDITIONAL INFORMATION
Typical method events for an LC-MS analysis using LockSpray:
LockSpray properties:
Acquire LockSpray – Apply correction: Lockmass correction applied in real time, i.e. lockmass data is centroid
Acquire LockSpray – Do not apply correction: Lockmass data is in continuum – typically set for proteomics application where data is to be processed using PLGS
No LockSpray: Acquisition will not switch to the LockSpray
Lockspray reference compound: Select the correct reference compound used for lockmass and correct mode of operation
id219495, MALDIG2SI, MLYNX, MLYNXV41, SUPMM, SYNG2SIHD