How to integrate Drift Time (DT) versus Intensity DriftScope data - WKB21641

Article number: 21641

OBJECTIVE or GOAL

Integrate Drift Time versus Intensity DriftScope data (DT vs. I).

ENVIRONMENT

  • DriftScope
  • MassLynx
  • SYNAPT HDMS
  • SYNAPT G2
  • SYNAPT G2-S
  • SYNAPT G2-Si

PROCEDURE

  1. Select the desired Drift data in the DriftScope and select the green check mark green check mark.PNG so that only the desired data shows.
  2. Select the File drop-down menu > Export to MassLynx > Retain DriftTime.
  3. Go to MassLynx > Chromatogram and select File > Open.
  4. Open the File Name_dt_0# file that was just created.
  5. Select Display >View and change the Axis Label Horizontal Axis to Scan.
  6. The Drift Time Chromatogram is ready to be integrated.
  7. Select Process > Integrate, or the Integration button Integration.PNG or process in TargetLynx.

ADDITIONAL INFORMATION