Processing error (-45) when processing MassLynx 4.1 data with MaxEnt1 - WKB21387
Article number: 21387
SYMPTOMS
- Processing error (-45) when processing data with MaxEnt1
- Rebooting the PC does not help
- Found: No antivirus software is installed
- Found: User is logged in to PC as local Administrator
ENVIRONMENT
- MassLynx 4.1
CAUSE
Corruption of the Masslynx.ini file
FIX or WORKAROUND
- Close MassLynx and all MassLynx-related windows.
- Browse to C:\Masslynx and find the Masslynx.ini file.
- Open the Masslynx.ini file in Notepad.
- Find the [SPeMaxEnt] section in the Masslynx.ini file.
- Delete the information in this section, leaving only the [SPeMaxEnt] section header.
- Save the Masslynx.ini file.
- Open MassLynx.