Error: Zp-axis move h/w fault - WKB208489
Article number: 208489
SYMPTOMS
- "Zp-axis move h/w fault" encountered when trying to inject sample
- The error is triggered by switching to ANSI 96-well 1-mL plate (WAT058957)
- The error does not persist when using a different plate configuration
- Plate configuration and vial format are correct
- Not resolved after Z axis calibration
- Same issue applies when the 48-well 2.mL plate is used instead of the ANSI 48-well 2-mL plate
ENVIRONMENT
- ACQUITY Sample Manager FTN
- ACQUITY Sample Manager FL
- ACQUITY Arc HPLC Sample Manager FTN-R
- Empower
- MassLynx
CAUSE
- The default Z value for the ANSI 96-well 1-mL plate size in the plate configuration is not suitable for Waters 1-mL 96-well plate (WAT058957).
- The default Z value for the 48-well 2-mL plate size in the plate configuration is not suitable for Waters ANSI 2-mL 48-well plate.
FIX or WORKAROUND
- Change the Z value from 310 to 425 (same as the ANSI 96-well 2-mL plate).
- Change the Z value accordingly for the 48-well 2-mL plate, in order for it to match the value of the ANSI 48-well 2-mL plate.
ADDITIONAL INFORMATION
id208489, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, MLYNX, MLYNXV41, SUP, SUPMM