FTN rotary tray misalignment - WKB201687
Article number: 201687
SYMPTOMS
- Misalignment of rotary tray
- Even if you select plate number 1, the plate is not in front of you but at an angle
ENVIRONMENT
- ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)
- ACQUITY UPLC H-Class
CAUSE
If the rotary tray is replaced with firmware 1.45 or lower, the tray position will shift.
FIX or WORKAROUND
- Upgrade firmware to 1.50 minimum.
- Perform rotary tray calibration.