How can I reduce the MassLynx data file size? - WKB20159
Article number: 20159
ENVIRONMENT
- MassLynx
- Long run reaching the 4 GB file size limit
ANSWER
- Have more intelligent MS method generation - If you know the retention time of your peaks collect data around those time points rather than the whole run.
- Collect data in centroid mode - this would have the greatest impact.
- Set up a solvent delay so the wasted time at the start of the run is not collected.
- Ensure the signal threshold has been set correctly.
- Increase the noise threshold.
- Increase the scan time, dwell time, interscan delay etc.