Cannot prime the ACQUITY UPLC Binary Solvent Manager because the option is dimmed in the Control menu in the ACQUITY Console - WKB200065
Article number: 200065
SYMPTOMS
- Cannot prime ACQUITY UPLC BSM
- Prime option in the Control menu in the ACQUITY Console is dimmed (unavailable)
- Cannot reset ACQUITY UPLC BMS because this option is also dimmed
- System status in MassLynx and ACQUITY Console says "ready"
- Inlet status in the MassLynx Inlet Editor shows green LED for READY and OK
ENVIRONMENT
- MassLynx
- ACQUITY UPLC Binary Solvent Manager
- ACQUITY Console
CAUSE
ACQUITY UPLC Instrument Control Software is not running properly.
FIX or WORKAROUND
- Close MassLynx and the ACQUITY Console.
- Run StopAcquityProcesses.cmd from the desktop icon or from C:\Program file (x86)\Waters Instruments\bin.
- Restart MassLynx.
- Open the ACQUITY Console.
ADDITIONAL INFORMATION
id200065, MLYNX, MLYNXV41, SUPMM, UPBINARY, UPBSM+, UPIBSM, UPIBSM+, UPPBSM