Cannot find inlet methods in MassLynx - WKB197389

Article number: 197389

SYMPTOMS

  • Cannot find inlet methods
  • Find multiple inlets listed vertically (odd), but cannot find methods in correct place
  • Saved a test method and it has only .ftn file

ENVIRONMENT

  • MassLynx

CAUSE

Corrupted inlet files

FIX or WORKAROUND

  1. Close MassLynx.
  2.  Log in to the Windows Registry and delete the inlets folders in both HKEY_CURRENT_USER and HKEY_LOCAL_MACHINE folders.
  3. Reboot the PC (the inlet files are re-created).

ADDITIONAL INFORMATION