Centroid MS spectrum of isotope is not displayed on MassLynx spectra screen - WKB192472

Article number: 192472

SYMPTOMS

  • Isotope spectra other than the highest intensity spectrum are not displayed

ENVIRONMENT

  • MassLynx

CAUSE

In the View settings for the spectra window, 60% is specified for “Data threshold%”.

FIX or WORKAROUND

Specify 0% for "Data threshold%".

ADDITIONAL INFORMATION