2695 does not communicate to the inlet method - WKB1846

Article number: 1846

SYMPTOMS

  • Download button (stack of papers) is unavailable
  • The pump on icon does not work—it is active, but when selected, the pump does not turn on
  • Remote icon is active
  • If taken out of remote, the front panel on the 2695 works correctly; flow will turn on
  • Hand icon is active—when you click on it, it says there is a method running, do you want to stop it?
  • If hand icon is stopped, still no communication
  • Ready and OK icons in the inlet method are green
  • No error messages appear
  • New inlet method does not solve the issue
  • New project does not solve the issue
  • The inlet method naming scheme is in the correct alphanumeric format
  • Reconfiguring inlet to Shimadzu and back to Alliance does not resolve the issue

ENVIRONMENT

  • MassLynx 4.1
  • 2695/Alliance

CAUSE

MassLynx software is corrupt (most likely cause), or the PDA must be reset.

FIX or WORKAROUND

  1. Close MassLynx.
  2. Run repair from add/remove programs under MassLynx (not tested/validated, so it isn't recommended, but it did work for this instance).
  3. Open MassLynx and reset PDA communication inside of the inlet method.

ADDITIONAL INFORMATION

 