2695 does not communicate to the inlet method - WKB1846
Article number: 1846
SYMPTOMS
- Download button (stack of papers) is unavailable
- The pump on icon does not work—it is active, but when selected, the pump does not turn on
- Remote icon is active
- If taken out of remote, the front panel on the 2695 works correctly; flow will turn on
- Hand icon is active—when you click on it, it says there is a method running, do you want to stop it?
- If hand icon is stopped, still no communication
- Ready and OK icons in the inlet method are green
- No error messages appear
- New inlet method does not solve the issue
- New project does not solve the issue
- The inlet method naming scheme is in the correct alphanumeric format
- Reconfiguring inlet to Shimadzu and back to Alliance does not resolve the issue
ENVIRONMENT
- MassLynx 4.1
- 2695/Alliance
CAUSE
MassLynx software is corrupt (most likely cause), or the PDA must be reset.
FIX or WORKAROUND
- Close MassLynx.
- Run repair from add/remove programs under MassLynx (not tested/validated, so it isn't recommended, but it did work for this instance).
- Open MassLynx and reset PDA communication inside of the inlet method.
ADDITIONAL INFORMATION
id1846, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, A-10SSM, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB, comms, MLYNX, MLYNXV41, SUPMM