Sample list pauses with "Waters Generic Instrument COM Server Detector Method Failed to Open Raw Data Files" error - WKB1751

Article number: 1751

SYMPTOMS

  • Sample manager collects sample, moves to the injection port, and then the sample list pauses
  • "Waters Generic Instrument COM Server. Detector Method Failed to Open Raw Data Files" error is observed from the inlet editor
  • Creating a new project, method, and sample list did not resolve the problem alone

ENVIRONMENT

  • MassLynx 4.1
  • 2767 Sample Manager
  • ACQUITY UPLC

CAUSE

The sample list had 600 lines with sample and method names containing hyphens.

FIX or WORKAROUND

  1. Create a new sample list.
  2. Name all files and samples using only alphanumeric characters and the underscore.

ADDITIONAL INFORMATION

 