"Wrong ionization mode" error message in MassLynx when trying to run an MS method for API mode - WKB16857
Article number: 16857
SYMPTOMS
- Trying to run an assay in API mode
- Error "wrong ionization mode" when samples are queued up
- No error seen when running samples in ESI mode
ENVIRONMENT
- MassLynx, all versions and SCNs
- All MassLynx-controlled mass spectrometers
CAUSE
Mismatch between the ionization modes set in the MS Method and the Tune page
FIX or WORKAROUND
Change the ionization mode in the Tune page to match the mode required for the MS method set in the sample list.
ADDITIONAL INFORMATION
id16857, Electrospray, MLYNX, MLYNXV41, SUPMM