In MassLynx, the MS goes into Standby mode when running Start Up System from the console - WKB16599

Article number: 16599

SYMPTOMS

  • In IntelliStart, the MS flips from Ready/Operate to Standby mode when the Start Up System option is selected

ENVIRONMENT

  • MassLynx 4.1 SCN 862
  • ACQUITY UPLC I-Class
  • ICS 1.60

CAUSE

System Startup has the LAMP setting configured OFF.

FIX or WORKAROUND

  1. In the console, click on the System tree.
  2. Go to the Configure menu.
  3. Click on the choice called Start up system....
  4. On the Detectors tab, confirm that the LAMP setting is ON.

ADDITIONAL INFORMATION

 

 