How can Empower data be imported into MassLynx 4.1? - WKB16563

Article number: 16563

ENVIRONMENT

MassLynx 4.1 (all releases) with Databridge

ANSWER

  1. Export the channel from Empower to .cdf.
  2. Use Databridge to convert the .cdf file to a MassLynx raw data file.

ADDITIONAL INFORMATION

  • Databridge is installed with MassLynx 4.1. It is not installed with MassLynx 4.1 Security.
  • Ensure that there are no spaces in the file names.
  • Databridge can output only Centroid data, so Continuum data in Empower will be Centroided during the conversion.
  • PDA data cannot be converted.
  • Only an official Waters image of Windows (7 or 10) can be used.
  • Only an official Waters PC (such as a Lenovo P510, P520, or P720) can be used.