Skip to main content
Waters

Error message: Positive and negative resolution expiry - WKB1438

Article number: 1438

SYMPTOMS

  • System status: Not ready
  • Positive and negative resolution file expiry

ENVIRONMENT

  • MassLynx 
  • Empower 3
  • Xevo TQ-XS
  • ACQUITY TQD system
  • SQ Detector 2
  • Xevo TQ Absolute
  • Xevo TQ Absolute XR 
  • Xevo TQ MS
  • Xevo TQD
  • Xevo TQ-GC
  • Xevo TQ-S
  • Xevo TQ-S cronos
  • Xevo TQ-S micro

CAUSE

  • The Positive and negative resolution file created by Intellistart has expired.

FIX or WORKAROUND

  1. If system is bound by an SOP to check the resolution every X number of days, user will need to rerun the Intellistart Resolution Setup to clear the error.
  2. If system is not bound SOP to check the resolution on a schedule, the expiration check can be disabled:
    • Go to the MS Console
    • High light Intellistart in the menu tree on left
    • Go to Configure menu and select Intellistart Configuration
    • Uncheck Positive and Negative Resolution File expired

Intellistart_Config_1.JPG

Please noticed if all the about it unchecked, if one is missing the error will persist. 

ADDITIONAL INFORMATION

NOTE: This resolution is applicable to the expiration of resolution and calibration files as well as Service due for all quadrupole bases system controlled by Masslynx or Empower. For most systems the option is Configure > Intellistart configuration. For the QDa the option is Configured > Health Monitor configuration and may require that the Console be set in Service Mode. 

Intellistart is designed to alert the users when certain files have expired (resolution setup, mass calibration, tune file, etc.) or when the system needs to be PM'd (service due). These checks are meant for regulated labs whose SOP dictate that work cannot be done if these files are outside the number of allowable days. The checks bring up messages that restrict the system from running samples. The default is 180, commonly set to 365. If labs are not subject to such restrictions, then the expiration checks can be disabled.

To change the number of days. Highlight the line you wish to alter in console/configure/intellistart configuration/ and press properties  The page then shown will allow you to change the number of days from the default 180.

Not able to find a solution? Click here to request help.