For one analyte, there are no values in the response and concentration columns in TargetLynx - WKB12920
Article number: 12920
SYMPTOMS
- No values in the response and concentration columns for one analyte
- Peaks are integrated and identified
- Data for other analytes appears good
ENVIRONMENT
- MassLynx
- TargetLynx
CAUSE
The concentration of the internal standard for that one compound was not set to Fixed.
FIX or WORKAROUND
- Select the internal standard in the TargetLynx method.
- Change the Concentration of Standard: Level to Fixed.
- Save the method.
- Reprocess the data.
ADDITIONAL INFORMATION