No concentrations reported for some of the compounds in the Targetlynx results - WKB12375
Article number: 12375
SYMPTOMS
- No concentration reported from some compounds in the Targetlynx summary table.
- Method uses Absolute Response
- Found: Response column is summary is populated
- Found: Primary Detection Flag is bbI for compounds with no concentration reported.
ENVIRONMENT
- Masslynx 4.1
- Targetlynx 4.1
CAUSE
The compound is flagged as indeterminate because the response is a smaller value than the intercept of the calibration curve.
FIX or WORKAROUND
- There are two possible resolutions.
- Prepare a new calibration curve where the resulting intercept from the regression will be a smaller value.
- Change the Calibration Origin setting in Calibration Properties of the Targetlynx method to Forced setting the intercept to a value of zero.
ADDITIONAL INFORMATION
Note that in MassLynx 4.2, forcing the curve through the origin will not help.