Problem using Trend Micro OfficeScan antivirus software with MassLynx - WKB12352
Article number: 12352
SYMPTOMS
- Problems include:
- Cannot open the MassLynx inlet editor
- Cannot open the Tune page
- Tune page not responsive or options greyed out
- Run samples pane freezes when Start pressed "not responding" error message
- Communications losses with MassLynx
ENVIRONMENT
- MassLynx
- Trend Micro antivirus software
- Trend Micro Apex One Security Agent
CAUSE
Incompatibility between (unsupported antivirus software) Trend Micro configuration and MassLynx
FIX or WORKAROUND
In Trend Micro disable the following:
Application Control
Behavior Monitoring
Data Loss Prevention
Device Control
Endpoint Sensor
Firewall
Outbreak Prevention Policy
Predictive Machine Learning
Suspicious Connection Service
Vulnerability Protection
Consider disabling SmartScan