PDA or other chromatograms do not display automatically from the MassLynx sample list using the "Chromatograms to Display" function - WKB119364
Article number: 119364
SYMPTOMS
- When chromatogram is opened from the sample list, only the MS TIC function is displayed; the PDA chromatogram or other specified traces are not displayed
- Previously worked fine
- The parameters in Tools > Options > "Chromatograms to Display" are correctly filled in for the PDA (DAD TIC), mass chromatogram, and others
- The specified chromatograms open correctly on a different MassLynx PC with the same software
ENVIRONMENT
- MassLynx
CAUSE
Unknown - possible corruption in the software
FIX or WORKAROUND
Reinstall MassLynx.
As a workaround, manually add the required traces to the chromatogram window or use ChroTool.
ADDITIONAL INFORMATION
Note: Swapping out the MassLynx.ini file does not fix the issue.
A similar issue has been reported by a different customer with the display of the MS TIC.
They solved the problem by rebooting the workstation a few times.
Another customer reported that when displaying a data file, the UV trace was always displayed instead of the TIC. It previously worked fine.
Solved by re-installing MassLynx.
id119364, MLYNX, MLYNXV41, SUPMM