"Insufficient disk space" error in ACQUITY system status - WKB1105
Article number: 1105
SYMPTOMS
- ACQUITY system status: Insufficient Disk Space
ENVIRONMENT
- MassLynx 4.2
- ACQUITY MS Console
- IntelliStart
CAUSE
IntelliStart appears to be checking the hard drive and memory space on the EPC.
FIX or WORKAROUND
- Open the ACQUITY MS Console.
- Select IntelliStart under the mass spec in menu tree.
- Click Configure > IntelliStart Configuration.
- Clear the Disk Space option, and then click OK.
ADDITIONAL INFORMATION
IntelliStart monitors disk space as part of the workstation's parameters, and it is recommended to ensure sufficient disk space on the workstation's drives. Specifically, the operating system hard drive should have more than 20% free space, and the Windows C drive containing the operating system and program files must be less than 80% full