Unable to complete execution of startup: failure to load LC method - WKB108066
Article number: 108066
SYMPTOMS
- Unable to complete execution of startup; failure to load LC method
- System pauses in between batches when multiple batches are in the queue
ENVIRONMENT
- MassLynx 4.1 and 4.2
CAUSE
The startup LC method is not selected, or the method selected no longer exists.
This is a universal MassLynx setting.
FIX or WORKAROUND
- Select the correct startup method from the startup/shutdown menu.
- System startup is not recommended if multiple LC methods are needed or multiple batches are in the queue.
ADDITIONAL INFORMATION