How to subtract a blank baseline from chromatograms in a Sample Set when working with 3D data - Tip273

Article number: 232408

OBJECTIVE or GOAL

How to subtract a blank baseline from chromatograms in a Sample Set when working with 3D data for both PDA and MS data.
 

ENVIRONMENT

  • Empower
  • Empower Tip of the Week #273

PROCEDURE

STEP 1
Subtract a baseline collected over a 3D wavelength range by entering the label of the blank sample in the ‘PDA 3D Blank Subtraction’ field within the Method Set (figure 1).
Figure_1.png

STEP 2
Once the Sample Set is Processed, the blank subtracted from itself yields a straight line. All subsequent chromatograms will have a baseline which reflects the subtraction of the blank baseline (figure 2).
Figure_2.png
MS Data: This is useful if your QDa detector becomes contaminated by something like polyethylene glycol, which can interfere with the MS spectra in your sample. Using the MS 3D Subtraction in the Processing Method, you can subtract a background spectrum to eliminate the interference. Note: this only works when you interactively process data in the Review window. It does not work with background processing.

STEP 3
Start by looking at a TIC plot for a 3D MS channel in Review (figure 3).
Figure_3.png

STEP 4
Click the 3D Plot tool in the tool bar or select 3D Plot from the Window menu. The plot displays time on the x-axis, intensity on the y-axis and mass on the z-axis. Note: the high intensity at the low mass end highlighted in blue (figure 4).
Figure_4.png

STEP 5
On the MS 3D Channel tab in the Processing Method, we can enable the MS 3D Subtraction (figure 5). 
Let’s look at the parameters. 

  • Identify an area(s) of the baseline without peaks which can be used for the subtraction. Tick the Enable MS 3D Subtract box and enter a time range or ranges in the Subtract field. A time range is entered using the ‘:’ (colon) between times and a ‘;’ (semi-colon) separating more than one range. Empower will calculate an average spectrum based on the times entered which will be subtracted from everything else.
  • Specify a value which to multiply the spectrum by before performing the subtraction.
  • Specify a noise threshold which will remove data points with intensities below the threshold. This value is entered as a percentage of the base peak intensity.
  • Empower uses Peak Separation to combine data points within the range to produce a single mass. For example: if you specify a value of 2 then all masses from 299 to 301 are combined into a single mass of 300.

Figure_5.png

STEP 6
Return to the Main Window and reintegrate the chromatogram. From the 2D channels tab, select the channel which resulted from the subtraction and overlay it with the original TIC plot. Notice the change in the baseline and the peaks between 4.2 and 5 minutes (figure 6). 
Figure_6.png

STEP 7
Select the chromatogram with the baseline subtraction by highlighting it in the 2D channels tab and click the 3D Plot tool. Notice the intensity at the low mass range has been reduced (figure 7). 
Figure_7.png

STEP 8
Return to the Processing Method and enter a value for the Noise Threshold. Reintegrate the chromatogram in the Main Window. From the Spectra table under the Spectrum Review window, deselect all but one spectrum (figure 8). 
Figure_8.png

STEP 9 
Click the Spectrum Points tab to view each point of the spectrum with its mass and intensity. If you compare the point table before and after the use of the Noise Threshold you will observe a difference in the number of points (figure 9). 
Figure_9.png