Skip to main content
Waters

Working with Derivatives of UV Spectra and the Splining Function - Tip264

Article number: 227375

OBJECTIVE or GOAL

In this article for Empower Chromatography Data System (CDS), we will work with derivatives of UV spectra and also learn about the Splining function

ENVIRONMENT

  • Empower
  • Empower Tip of the Week #264

PROCEDURE

Derivatives

STEP 1

Start with the UV spectrum for Caffeine in Sample 1, the peak in this sample is spectrally pure (figure 1).
Figure_1.png

 

STEP 2

From the Spectrum Review menu, select ‘Keep Spectra When Switching Channels’ (figure 2).
Figure_2.png

 

STEP 3

Select the UV spectrum for Caffeine in Sample 2, the peak in this sample has an impurity as we saw in an earlier tip. The stack plot of the two spectra shows that the shape and lambda max appear to be the same (figure 3).

Figure_3.png

 

STEP 4

Empower can use derivatives of UV spectra to enhance the differences between them. In the Spectra table, select ‘Second’ in the ‘Deriv. Level’ field (figure 4).

Figure_4.png

 

STEP 5

Now the differences between the spectra are discernible and are due to the impurity under the Caffeine peak in Sample 2. Smoothing is automatically applied to the derivative of the spectra (figure 5).

Figure_5.png

 

Splining

STEP 6

Splining can enhance the display of a UV spectrum. We see the original UV spectrum for Caffeine with a digital resolution of 0.6nm (figure 6).

Figure_6.png

STEP 7

Select ‘Spline’ in the Spline field for the spectrum (figure 7).

Figure_7.png

 

STEP 8

The digital resolution is now 0.2 nm. The interpolated points are calculated by performing a cubic Spline operation (figure 8).

Figure_8.png

ADDITIONAL INFORMATION

Final Note: This can be done with either the Pro or QuickStart interface.

Not able to find a solution? Click here to request help.