Skip to main content
Waters

How to calculate resolution between peaks that are in different channels (2) - Tip261

Article number: 226694

OBJECTIVE or GOAL

This article discusses an alternative approach by creating a Custom Field to calculate resolution between the two peaks. This would be a Peak Custom Field using Intersample syntax. 

ENVIRONMENT

  • Empower
  • Empower Tip of the Week #261

PROCEDURE

STEP 1
We have a stack plot of the two channels of data, each with its own peak eluting at different retention times (figure 1).
Figure_1.png

STEP 2
Be sure to enable System Suitability in the Processing Method as we will need the ‘Width at Tangent for USP Resolution’ values for our Custom Field (figure 2).
Figure_2.png

STEP 3
The calculation for USP Resolution = 2(RT2-RT1)/(W1+W2), where RT is the ‘retention time’ and W is the ‘width at Tangent’ for USP Resolution (figure 3).
Figure_3.png

STEP 4
Intersample syntax is ‘label.injection.channel (field). (Refer to Empower Tip #93 for more details on creating intersample Custom Fields) (figure 4).
Figure_4.png

STEP 5
The properties of the Custom Field can be viewed (figure 5).
Figure_5.png

STEP 6
The result displaying the resolution between the two peaks in the two different channels is different from the value calculated in Tip #260. The baseline in the combined chromatogram is different from the baselines in the individual chromatograms resulting in different baseline noise, different peak starts/stops and different peak widths. The key is to pick one approach for consistency in the calculations (figure 6).
Figure_6.png
 

ADDITIONAL INFORMATION

Final Note: This can be done with either the Pro or QuickStart interface. 

Not able to find a solution? Click here to request help.