PDA Processing Method: Spectral Contrast Parameters - Tip254

Article number: 224337

OBJECTIVE or GOAL

This article discusses the Processing Method for the Photodiode Array (PDA) Detector, starting with the Spectral Contrast parameters

ENVIRONMENT

  • Empower
  • Photodiode Array Detector (PDA)
  • Empower Tip of the Week #254

PROCEDURE

STEP 1
When creating a new Processing Method, be sure to select ‘PDA’ and click the ‘Purity’ tab. The default Spectral Contrast values appear (figure 1).


 Figure_1.png

 

STEP 2
The ‘Wavelength Limit’ should be set to start above the UV cutoff of the mobile phase and end at a value that includes all absorbances. Start with ‘MAX Plot’ and look at the UV spectra for the peaks of interest (figure 2). 


Figure_2.png

 

STEP 3
Enter the ‘Wavelength Limit’ in the ‘Start’ and ‘Stop’ (figure 3).


Figure_3.png

 

STEP 4
The ‘Noise Interval’ in minutes should be determined from ‘MAX Plot’ as well. This interval should be 2–4 peaks widths (at half height) long. Remember to multiply the ‘interval’ by the ‘Sampling Rate, from the Instrument Method, to calculate the number of data points in the interval, as it must contain a minimum of 12 points (figure 4).


Figure_4.png

 

STEP 5
Zoom in on an area of the baseline that does not have any significant absorbance (figure 5).


Figure_5.png

 

STEP 6
Enter the region determined from Step 5 into the ‘Noise Interval’ field (figure 6).


Figure_6.png

 

STEP 7
Verify the ‘Noise Interval’ by selecting ‘Display Noise Spectrum’ from the ‘Spectrum Review’ menu. Each value plotted is the ‘root-mean-square noise’ at each wavelength in the spectrum. The shape of the noise spectrum should be random and not resemble the spectra from the peaks of interest. The noise spectrum is chemistry dependent (figure 7).


Figure_7.png

 

STEP 8
If the ‘Noise Interval’ contains less than 12 spectra, Peak Purity will NOT be calculated, and you will see messages in the Message Center indicating that the Noise Interval is invalid and does not contain sufficient spectra (figure 8).


Figure_8.png


 

ADDITIONAL INFORMATION

Final Note: This can be done with either the Pro or QuickStart interface.

Not able to find a solution? Click here to request help.