Skip to main content
Waters

Empower 3: How to remove/add RIDs from the Agilent LC Chromatographic System Configuration - WKB99276

Article number: 99276

OBJECTIVE or GOAL

  • Remove/add RIDs from the Agilent LC Chromatographic System Configuration

ENVIRONMENT

  • Empower 3
  • Agilent LC
  • ICF3.0 and later

PROCEDURE

1.If there is an existing system for the Agilent stack, right-click on the system and select Takeoff Line before opening the Agilent Prconfiguration.

2.Open the Agilent PreConfiguration window

On the Configuration Manager window, click the "Tools" menu > "Agilent PreConfiguration" .

clipboard_e4aab9234ad7b278c0eed34c7fdda5157.png

The Configuration Directory window opens.

3. Select the relevant devices on the Configuration Directory window.

Enter the LAC/E name or LAC/E IP address to which the relevant LC is connected in "IP address/host name" on the Configuration Directory window, and then click "Connect".


clipboard_ef836cc251010d201082cf9f4a5607ef1.png

Click the displayed LC icon.
On this screen, the system is identified by the IP address of the LC.

4. Click OK after modifying the configuration on the Configuration Editor window.

Make sure that all LC modules used in this measurement, including the RI detector, are powered on.
Select the relevant LC on the Configuration Directory window and click "Configurations".

clipboard_e9ee82329ffbbf85825d269fe7ef418c2.png

Click on Clear button to remove the device displayed in the list on the right side of the Configuration Editor window.

clipboard_edaa5c5fdc66c3ad39b836c57ca52a830.png

After making the list on the right side of the Configuration Editor window blank, select the LC system name to be used in the list on the left side and click "Automatic Configuration".

clipboard_e203df5ee66658148e3a928154205be71.png

Select IP address, enter IP address and click Ok

clipboard_eca219d4da5078b9b845d094db0583795.png

Information about the powered-on LC module is automatically loaded and displayed in the list on the right.

clipboard_ecf7d6b2a4970ccc9f61528bf1b2b0ca4.png

Confirm that the RI detector has been added to the list on the right.
If necessary, set the pressure unit of the pump and the temperature control of the autosampler.
Click OK to close the widnow

* Be sure to click "OK" to close the Configuration Editor window. If you close with the X mark, the settings will not be saved.

 

ADDITIONAL INFORMATION

 

id99276, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP

Not able to find a solution? Click here to request help.