Are SampleWeight and Dilution used in the calculation of Amount in Empower? - WKB96394
ENVIRONMENT
Empower 3
ANSWER
Yes.
ADDITIONAL INFORMATION
From Empower Help:
Dilution
Factor by which Empower software (during calibration) divides the calculated amount or concentration of each standard component. During quantitation, Empower software multiplies the amount (or concentration) read from the calibration curve (of the standard components) by the Dilution value to calculate amounts and concentrations for each unknown sample.
SampleWeight
Factor by which Empower software (during calibration) multiplies the calculated amount or concentration of each standard component. During quantitation, Empower software divides the amount (or concentration) read from the calibration curve of the standard components by the SampleWeight value to calculate amounts and concentrations for each unknown sample.