Agilent LC or GC is not recognized in Empower node properties after ICF upgrade - WKB94990
Article number: 94990
SYMPTOMS
- After upgrading ICF from 3.0 to any version or 3.1 to any version or 3.2 to any version, Agilent LC or GC shows "NO" as instrument status in node properties
- After upgrade, either "server busy" or "instrument failure" may occur while starting a sample set
ENVIRONMENT
- Empower 3
- Agilent Instrument Control Framework
- Agilent LC
- Agilent GC
CAUSE
The software deletes the configuration details during upgrade
FIX or WORKAROUND
- Go to Empower Node Properties and remove the instrument from "Config DHCP".
- Add the instrument again with MAC address and unique instrument name.
- Scan the instruments and ensure that Agilent LC or GC shows "YES" as instrument status.
ADDITIONAL INFORMATION
CRI-1383