Only the instrument with the LAN cable plugged into it can be seen in Empower for an Agilent Infinity II System - WKB9468
Article number: 9468
SYMPTOMS
- Only the module with the LAN cable plugged into it can be seen
- When the InstantPilot is connected, all of the modules appear
ENVIRONMENT
- Empower 3
- Agilent Infinity II LC
- G7165A, G7129A, G7112B, and G4208A InstantPilot
- On/off status of the third-party mode on the InstantPilot makes no difference
CAUSE
The Agilent LC stack is in Resident Firmware mode.
FIX or WORKAROUND
Set the firmware mode back to "Main".
ADDITIONAL INFORMATION
If not able to set the firmware mode back to Main, see the following link for an Agilent cold start reset: