2475 fluorescence detector does not appear in the Empower "Run Samples" control panel - WKB9341
Article number: 9341
SYMPTOMS
- The 2475 fluorescence detector does not appear in the Empower "Run Samples" control panel
- No communication issues
- Able to collect data and run normally
ENVIRONMENT
- 2475 Fluorescence (FLR) detector
- ICS 1.01
- Firmware 2.03
CAUSE
The control panel in Empower Run Samples is not available with ICS 1.01.
FIX or WORKAROUND
- Upgrade to Waters Driver Pack 2016 Release 1.
Note : Starting with ICS 1.65 and firmware 2.10, which is Waters Driver Pack 2016 Release 1, the control panel appears in the Run Samples window.
ADDITIONAL INFORMATION
id9341, 2475, 2475B, A-10FL, comms