Empower backup is missing EXP, INF, CKS and integrity files - WKB9320
Article number: 9320
SYMPTOMS
- No errors involving the backup of the project except failed to read raw data file messages
- Projects are missing the integrity.txt file and EXP , INF and CKS files in the backup folder
- On the Integrity tab of the project properties, integrity.txt files are missing from the project, even if the project is backed up every week
- The tablespace in Oracle has enough free space for the project
- Error Message "No restorable directories found in the root directory. Restorable directories must contain an EXP and an INF file." during back-up (see screenshot)
ENVIRONMENT
- Empower 1
- Empower 2
- Empower 3
CAUSE
The files are not able to be moved with the operating system. Raw data files are unable to be copied successfully, causing the backup to fail prior to the .exp file being written.
FIX or WORKAROUND
- Check for, and free up, disk space on the Empower client machine.
- Back up projects through Empower.
ADDITIONAL INFORMATION
29Nov2024:
Empower is designed to restore a complete backup project.
Empower is built on a Oracle database. Thus, the project is moved in & out, as a part of the database.
The project restore will never work in Oracle database if you selected some files from various back-up folders and put into into new folder.
This only work for flat file format, which the architecture is commonly built in other CDS.