Missing "Cloned Project" system audit trail event for cloned parent project in Empower - WKB90904

Article number: 90904

SYMPTOMS

  • Parent project and child projects exist
  • Cloned Project events are present in the System Audit Trail for child projects
  • Cloned Project event is not present in the System Audit Trail for parent project

ENVIRONMENT

  • Empower 3

CAUSE

The cloning process was interrupted. Due to the order of project creation and the current order of writes to the System Audit Trail, the parent project is created first, but the write to the System Audit Trail for this event in written last.

FIX or WORKAROUND

Ensure that the clone process is not interrupted.

ADDITIONAL INFORMATION

This issue is documented as PCS 53097 and CRI-1762. This behavior is corrected in Empower 3.8.0