How do you add the MW markers and slice in GPC using Empower? - WKB89361

Article number: 89361

ENVIRONMENT

  • Empower 3
  • Gel permeation chromatography (GPC)
  • MW markers

ANSWER

  1. Open the GPC Processing method.
  2. Select Broad.
  3. Click on the Slicing tab.
  4. Scroll to the right and specify the MW markers.
  5. Save the processing method.
  6. Process and save results.
  7. Highlight the results and bring them to Preview Publisher.
  8. Open using Default Report with Slice Table.
  9. Click Close so that you can edit.
  10. Click inside the table, right-click, and choose GPC Distribution Table Properties.
  11. Click the Slice tab > Slice by MW markers > Apply > Close.
  12. Open the Preview window. 

ADDITIONAL INFORMATION

The options for displaying slices do not require that the data be calculated differently. Empower slices at every data point, but the slice table can be customized as to how the slices are displayed.