Empower Users require Create Node Privilege in order to log in via Citrix - WKB89342
Article number: 89342
SYMPTOMS
- Error - Your user profile does not allow you to create a new node. Please contact your Administrator.
ENVIRONMENT
- Citrix
- Empower
- SR3
- FR5
- SR4
CAUSE
When a user logs in to Empower via Citrix, a new Citrix Node Type is created in Empower.
Without the Create Node privilege, users are presented with the error message "Your user profile does not allow you to create a new node. Please contact your Administrator."
May also occur on a Workstation.
FIX or WORKAROUND
- User requires the Create Node privilege.
- Alter Node and Delete Node privileges should be restricted.
- Access to Nodes can be restricted by groups in configuration manager. User who is not a member of a group with access cannot see the node listed or access the node properties from instruments (systems) they can access.
Note: This issue was addressed in EMP 3.8.0
ADDITIONAL INFORMATION
CRI-309 - was created for this issue.
Any user who does not have Create privileges cannot log into Empower for the first time on Citrix.
Empower 3 SR3 introduced more granularity of the node privileges.
|
Create Node |
Add nodes to your Empower environment. |
|
Alter Node |
Change properties of nodes within your Empower environment. |
|
Delete Node |
Delete nodes from your Empower environment. |