Inconsistent extraction error when using Agilent 7697 Headspace and Empower - WKB8872

Article number: 8872

SYMPTOMS

  • An Agilent 7697 "inconsistent extraction" error message appears in Empower
  • "HS7697A: Inconsistent extractions. No. of injections in sample set method should match the no. of extractions in instrument method."

ENVIRONMENT

  • Empower
  • Agilent 6890
  • Agilent 7890
  • Agilent 7697 Headspace

CAUSE

A GC 7890 or 6890 configuration issue.
Auxiliary Temp is not set properly. 

FIX or WORKAROUND

Correct the Auxiliary parameter for Agilent GC 6890 or 7890 (even if the error message does not seem to be related).

ADDITIONAL INFORMATION

The same error message may appear if the GC Cycle is incorrect.

See article WKB97285 on Headspace Sample preparation